Metal Coating Procedure
Equipment is Hummer 6.2 Sputter Coater, Anatech USA and Thickness Monitor Model TM-350, Maxtek Inc.
| 1. Check to make sure the correct target is installed. | ||||||
| 1.1. Gold is most commonly used. | ||||||
| 1.2. Pt/Pd is common for FESEM. | ||||||
| 1.3. If the target needs to be changed, contact the support staff. | ||||||
| 2. Lift the top of the bell jar to rest on the pipe behind it. | VIEW | |||||
| 3. Carefully remove the bell jar. | VIEW | |||||
| 4. Load specimens. | ||||||
| 5. Replace the bell jar. | ||||||
| 6. Lower the lid and check for fit. | VIEW | |||||
| 7. Turn on POWER, and hold the lid down until vacuum display no longer reads 1. | VIEW | |||||
| 8. Allow vacuum to pump to 100. Switch GAS to ON. | VIEW | |||||
| 9. Adjust the flow knob (located behind bell jar) by turning it in the positive direction (counterclockwise) until the Vac=200. | VIEW | |||||
| 10. Switch GAS to OFF until Vac=100. | ||||||
| 11. Switch GAS to ON up to 200 again. | ||||||
| 12. Switch GAS to ON, but this time adjust flow knob by turning it in the negative direction (clockwise) until Vac goes down to ~80. | ||||||
| 13. Turn on THICKNESS monitor. | VIEW | |||||
| 14. Press STOP then START. (The shutter light will turn red.) | VIEW | |||||
| 15. Hold the THICKNESS ENDPOINT button down to see thickness. | VIEW | |||||
| 15.1. For this display, 50 nm = 0.050 kA, and 5 nm = 0.005 kA. | ||||||
| 15.2. If the thickness needs to be adjusted, push the Thickness Endpoint button and press the up or down arrows to reach your desired thickness. | VIEW | |||||
| 16. Check that the VOLTAGE knob is set to zero. | VIEW | |||||
| 17. Switch VOLTAGE CONTROL to ON. | VIEW | |||||
| 18. Turn VOLTAGE knob up to 8 or less in order to obtain the PLASMA DISCHARGE CURRENT = 15 mA. | VIEW | |||||
| 18.1. For JEOL JSM-6400 SEM specimens use 15 mA. | ||||||
| 18.2. For Hitachi S-4700 FE-SEM specimens use 5 mA. | ||||||
| 19. When thickness endpoint is reached the current gage will automatically drop down to zero. | VIEW | |||||
| 20. Turn off THICKNESS monitor. | ||||||
| 21. Lower VOLTAGE knob to zero. | ||||||
| 22. Switch VOLTAGE CONTROL to OFF. | ||||||
| 23. Switch GAS to OFF. | ||||||
| 24. Switch off POWER. | ||||||
| 25. Let the air bleed and then remove specimens. | ||||||
| 26. Replace bell jar and lid. | ||||||