2.1. Log onto computer and load dQant32 software. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
2.2 Hit [E][S][P] icons in the dQant32 program (element, standard, points). |
|
|
| 2.3. Load the sample and start up the SEM. See Operating Procedures under JEOL JSM-6400 SEM. |
|
|
| 2.4. kV = 20, WD = 39, Tilt = 0. |
|
|
| 2.5. Adjust using the SE IMAGE BRIGHTNESS knob, SE IMAGE CONTRAST knob and MAGNIFICATION knob. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 2.6. Turn on the picoammeter. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 2.7. Push the blue PCD (Probe Current detector) button (lights up) on the control console to insert the probe current detector into the beam path. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 2.8. Press the [EOS] key on SEMs keyboard to launch a menu on the right SEM screen. Set beam current to 20 nA on the picoammeter using the CL. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 2.9. Set aperture (objective aperture assembly) to 1. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 2.10. Set CL coarse = 7, CL fine = ~55. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 2.11. Readjust CL coarse until the picoammeter reads 20 nA. When finished, turn off blue PCD button (light goes off). |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 2.12. Over the first ½ hour, the beam current will drop. To correct, adjust the CL fine to a smaller number. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 2.13. Flip the PCD switch from internal to external. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 2.14. Use the upper left black button on the joystick to move the PCD in and out of the beam path. Only use this button to toggle between in and out from now on. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
 |
 |
 |
 |
 |
 |
 |
| 3.1. Crank the EDS detector to the "in" position. See the NOTE in step 1.1 about elemental analysis before proceeding. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 3.2. In the "Standard Table: Master", locate the standards you will use. To highlight these, hold [Ctrl] and select your required standards, then release [Ctrl] and right click the mouse. Hit "Show Selected Records" to show only these selected standards in the Standard Table window. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 3.3. Use your standard map to locate these standards in the SEM. (As an aid in locating the standard, from the "dSspec" window, select menu "Stage > Specimen Holder Map".) Click on the stage map graphic where you want to move, then hold [Ctrl] and left click on the mouse at the same time. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 3.4. Highlight your standard name in the "Standard Table: Master". Click the "Set" button at the bottom of the table. A "Setup Standard" dialogue will appear. Click "Store". |
|
|
| 3.5. Set and save the stage position for standards and Benoite (MAC), or Willemite (ARL). These are the fluorescing samples needed later. Refer to the STD map. Use the EDS to ensure that the elements agree with the map. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
 |
 3.5.1. If in doubt about the standard's composition, choose Spirit software, click "erase", and "acquire". |
 |
 |
 |
3.5.2. Adjust the beam current so EDS dead time = 25-30%. |
 |
 |
 |
3.5.3. Stop acquiring, label peaks, go back to dQant32 screen. Go to the Standard Table and highlight your standard name. Click "set" and "store". Repeat above steps until all standards positions have been updated. |
 |
 |
| 3.6. Select "Point Table" in dQant32 screen. Select menu item "Tables > Add New Table". Name it. |
|
|
| 3.7. Move to the sample using "Specimen Holder Map". Save the stage position for your general work area in the Point Table. |
|
|
 |
3.7.1. Place the cursor in the first line of the new Point Table. |
 |
 |
 |
3.7.2. Click "set" and "store". |
 |
 |
 |
3.7.3. Repeat for all samples. |
 |
 |
| 3.8. NOTE: The software will support a Quantitative Line Transect. See support staff for instructions. |
|
|
 |
 |
 |
 |
 |
 |
 |
| 4.1. Flip the PCD to "in" and check for changes in beam current. Adjust the fine CL until 20 nA of beam current is displayed on the picoammeter. |
|
|
| 4.2. After the stage has moved, flip the PCD back to "out". If the screen is black, the PCD is "in", and the beam current will drop when its toggled. |
|
|
| 4.3. Focus in on a standard that is fluorescent. (Select "Standard Table" on dQant32 screen, select either Benoite or Willemite line, and click "Get".) |
|
|
| 4.4. Carefully focus the image on the SEM using the objective lens. |
|
|
| 4.5. Tilt the Optical Microscope (OM) down by grasping and pulling. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 4.6. Move the OM to the "in" position. You should be focused in on the fluorescing sample. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
 |
4.6.1. NOTE: If at any time your standard is not visible, or you can not find it, the OM must be moved back out. |
 |
 |
 |
4.6.2. Magnification must be 50x or greater. |
 |
 |
| 4.7. Press [PF6] to turn off the SEM scan mode. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 4.8. Make sure the PCD button is "out" on the joystick control. |
|
|
| 4.9. Locate the illumination from the beam through the OM. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
 |
4.9.1. Take the cap off. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
 |
4.9.2. Turn the LCD screen on. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
 |
4.9.3. Pull the knob out (on left side). When the knob is out, you can view through the objective lens. When the knob is in, you can view the image on the LCD instead. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
 |
4.9.4. Adjust OM using the locknut until the beam is in the center of the optical field. |
 |
 |
 |
4.9.5. Use the fine Z-axis The fine Z-axis stage translation is used to make fine changes in the distance between the specimen surface and the objective lens. SEM stage. control on the SEM to focus the illumination dot in the OM. |
 |
 |
 |
4.9.6. Push the knob back in. The LCD screen should show the illuminated dot. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
 |
4.9.7. Push the green [ON] button on the illumination console. The dot should be almost centered on the screen. |
 |
 |
 |
4.9.8. Refocus the image features (not the dot) with the Z control. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
 |
4.9.9. Use the objective lens (PROBE CURRENT knob on control panel 1) knob to produce the smallest dot possible. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 4.10. Pull the knob out (on the left side). |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 4.11. Move the OM crosshairs to coincide with the beam position on the fluorescing sample using the two knobs on the eyepiece |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 4.12. Flip the objective lens polarity switch on the control console twice (this removes hysteresis from the objective lens), then refocus with the objective lens and adjust crosshairs. Do not touch the objective lens setting after this point and do not change the crosshair position. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 4.13. Toggle the PCD in. Double check the beam current and adjust it back to 20 nA if necessary. |
|
|
 |
 |
 |
 |
 |
 |
 |
| 6.1. Go to "Your named Element Table". Select the desired element and click "Get Std". Check the OM focus and adjust with fine Z-axis control. |
|
|
| 6.2. Hit "CSearch" (coarse search) button. This drives the spectrometer to a within a few millimeters of the peak location and opens the Probe Current detector. |
|
|
| 6.3. In the peak search window you will see a representation of the counts. |
|
|
| 6.4. You should see a Gaussian shaped peak. If not, follow the steps in the Troubleshooting topic in JEOL JSM-6400 SEM. If that does not work, contact support staff. |
|
|
| 6.5. You may see error boxes appear. Hit "continue" until the error boxes go away. |
|
|
| 6.6. Coarse search is already in place, so now go to fine search by clicking the "FSearch" button. |
|
|
| 6.7. Now a higher resolution Gaussian peak is collected. |
|
|
| 6.8. Click "PkCalib". When the stage moves, its called a "jog" to relieve gear lash. |
|
|
| 6.9. "Focus Standard" dialog appears. Hit close. |
|
|
| 6.10. The "dSspec" window appears. Counts will be displayed. You will hear the detector moving. When the calibration is finished the new standard values will be stored, and displayed in "Your named Element Table". |
|
|
| 6.11. Repeat three times by hitting "PkCalib" 3 times. |
|
|
 |
6.11.1. Each time you hit "PkCalib", hit "close" on the "Focus Standard" dialogue afterwards. |
 |
 |
 |
6.11.2. Watch the peaks collect in "dSspec". |
 |
 |
| 6.12. Go to "Element"on the dQant32 menu and select "Average Calibration File…". The bottom entries are the most recent. |
|
|
| 6.13. Make sure the last three calibrations are similar under "net peak". |
|
|
| 6.14. Highlight those three rows (hold "Ctrl" while selecting). Hit "Apply Average" then "Close". |
|
|
| 6.15. Repeat steps 6.1 - 6.14 for each element that you want to measure. |
|
|
 |
 |
 |
 |
 |
 |
 |
| 7.1. Now set up the quant file. Go to the "Quant" menu and choose "Setup…". |
|
|
| 7.2. Hit "New Setup" button. |
|
|
| 7.3. In the "Setup File" field, replace the "New File" path name with your new file name. |
|
|
| 7.4. In the "Title" field add the title of the project. |
|
|
| 7.5. Input 20 to the keV field. |
|
|
| 7.6. In the "Dataset Folder" field, highlight the previous file name at the end of the path and replace it with your own file name. A dialog will tell you that the folder does not exist. Click "Yes" to create it. Alternatively, you can open "My Computer" on the desktop and create a folder, then type that path instead into the "Dataset Folder" field. |
|
|
| 7.7. Enter a unique name in the "Dataset File" field. |
|
|
| 7.8. Click "From Ele Table". It imports data from "Your named Element Table" to the "Quant Setup" table. The "Reference" column in "Quant Setup" should match the "Name" column in "Your named Element Table". Hit "Close". Click "Yes" to save changes made to the Quant Setup. |
|
|
 |
 |
 |
 |
 |
 |
 |
| 8.1. Go back to dQant32 "Your named Point Table", highlight the sample stage position, and click "get" (stage moves automatically). This moves to the position on the sample that you previously saved. |
|
|
| 8.2. Press [PF7] to turn on SEM scan mode. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 8.3. Check that the PCD is out. |
|
|
| 8.4. Lower the magnification. |
|
|
| 8.5. Adjust contrast and brightness. |
|
|
| 8.6. Reposition the sample to your desired analysis area using the stage control. Store the position in dQant32 points table. Be sure to write down which numbers correspond to which analysis points. (A line scan can be set up using this points table.) |
|
|
 |
8.6.1. Single point: Move stage to the desired location. All positioning will be done with the joystick while viewing the OM screen.
(If necessary, in the dSspec window, under the "Joystick" menu, select "Orientation Reversed" for easier control. Control will then be reversed in the SEM screen.)
Refocus with fine Z-axis. In "Your named Point Table", select the corresponding point number. Hit "Set". "Setup Point" dialog appears. Hit "4 - Store" and "Close". |
 |
 |
 |
8.6.2. Multiple points: Make a note of stage coordinates for your other specimens. Repeat step 8.6.1 above multiple times, once for each desired location. |
 |
 |
 |
8.6.3. Line of points: Repeat step 8.6.1. Reposition the stage to the location at the far end point for your line. In "Your named Point Table", select the second point number. Hit "Set". "Setup Point" dialog appears. Hit "4 - Store" and "Close". Then click "Calc. Line". The "Line Scan" dialog appears. Choose the number of points in your line OR the distance between points. NOTE: If you require evenly-measured spacings between points, it is easiest to move only in the X or Y directions. |
 |
 |
| 8.7. Press [PF6], view the sample through the OM, and focus with the fine Z-axis. Toggle the PCD in. Adjust beam current to 20 nA using the condenser lens for fine control. Press [PF7]. |
|
|
 |
 |
 |
 |
 |
 |
 |