| 1. TIP: These images have the best quality when the working distance is small or the CL setting is high (4 or 5). |
|
|
 |
1.1. Example 1: If the WD = 39, then the CL coarse = 4-6. |
 |
 |
 |
1.2. Example 2: If the WD = 15 or smaller, the CL coarse = 9-10. |
 |
 |
| 2. Log onto the computer, load the sample, and turn on the microscope. See Operating Procedures under JEOL JSM-6400 SEM. |
|
|
| 3. Set the aperture to 1. |
|
|
| 4. Locate the Backscatter Detector Switch (BSE preamplifier) on the column table. Set the switch to "SL". |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 5. Adjust the focus and the SE IMAGE BRIGHTNESS knob and SE IMAGE CONTRAST knob. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 6. Make your initial settings. |
|
|
 |
6.1. Control Panel 2, BE IMAGE CONTRAST knobs, set the "Coarse" knob to 4. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
 |
6.2. Control Panel 2, BE IMAGE CONTRAST knobs, turn the "Fine" knob fully clockwise. |
 |
 |
 |
6.3. Control Panel 2, BE IMAGE BRIGHTNESS knobs, turn the "Coarse" knob fully clockwise. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 7. Press the IMAGE SELECTOR COMP button to display a compositional image. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 8. Press the SCAN SPEED SLOW button. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 9. If there is no image, that means the brightness is too high. |
|
|
 |
9.1. Control Panel 2, BE IMAGE BRIGHTNESS knobs, turn the "Coarse" knob counterclockwise. |
 |
 |
 |
9.2. If the image is still too bright, for the BE IMAGE BRIGHTNESS knobs turn the "Coarse" knob fully clockwise again. For the BE IMAGE CONTRAST knobs, set the "Coarse" knob down one (to 3). |
 |
 |
 |
9.3. Repeat steps 9.1 and 9.2 until an image appears. |
 |
 |
 |
9.4. If the image is still not visible, use the CL coarse knob to lower the brightness. |
 |
 |
| 10. When the image is visible, use the DISPLAY WFM button and adjust the fine BE IMAGE BRIGHTNESS knobs and BE IMAGE CONTRAST knobs. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 11. Acquire and save images. See Acquiring a Digital Image in Basic Imaging under JEOL JSM-6400 SEM. |
|
|
| 12. When you are done backscatter imaging, return all settings to the original positions from step 6 in Backscatter. |
|
|
 |
 |
 |
 |
 |
 |
 |