Microanalysis
X-Ray Mapping
NOTE: The current drops quickly over the first 30 minutes of use. Wait until the current is stable before X-ray mapping.
| 1. Select Mode > EDS Image. | VIEW | |||||
NOTE: For the low resolution test collection, use a dwell of 5 ms. For any higher resolution maps, a dwell time between 5 and 20 ms is suitable. In the EDX Image Setup window, choose Edit Setup under EDX Image Processing. |
VIEW | |||||
| 3. Select elements by clicking in the element's box. | VIEW | |||||
| 4. To assign a different color to each element, right-click on the element > select Edit > Set Color > and choose a color. Click OK. | VIEW | |||||
| 5. Select OK in the EDX Image Setup menu once all of your choices have been made. | ||||||
| 6. Choose the correct magnification and kV in Options. NOTE: If the correct magnification is not in the preset menu, you will be able to manually enter the correct magnification in the Step 7 after selecting the camera icon. | VIEW | |||||
| 7. Perform a quick acquisition first by choosing a low resolution map of 320x240 in EDX Image. | VIEW | |||||
| VIEW | ||||||
| 9. Once the acquisition has begun, you can still edit the collection in live time by right-clicking in the image area and selecting ROI Processing. A menu with the same functions as the Setup menu will appear. Use the directions in Step 3 to edit. | VIEW | |||||
| 10. If the map is satisfactory, perform again using a high resolution. | ||||||
| 11. Save the map in your directory in the “Temporary User Files” folder. | ||||||