Imaging
Beam Current
NOTE: You will periodically need to re-set the beam current. To do so, click on the Set icon in the Accelerating Voltage window. Check to ensure the beam current is around 10 ľA at 15 minutes into your session and again at 45 minutes into your session.
Beam current (Ie) becomes more stable over time as contamination is spread evenly over the surface of the emission tip. |
Alignment and Stigmation
The following demos are used in Operation Step 8.
![]() |
![]() |
Operation
| 1. If you are the first user of the day, then peform the flashing procedure. See Flashing under Operating Procedures in Hitachi S-4700 FE-SEM. If you do not need to flash, select an operating voltage by clicking in the top right Accelerating Voltage window. | VIEW | |||||
| 2. An HV Control dialogue box will appear. Select the desired accelerating voltage. Click ON. | VIEW | |||||
| VIEW | ||||||
| 4. Once the operating conditions have been selected, roughly focus the image. | ||||||
| 5. To find an area of interest, select Low Magnification mode. Do this by clicking on the H/L icon to switch between high and low magnification. | VIEW | |||||
| VIEW | ||||||
| 7. Move to the area you want to image and focus. | ||||||
| VIEW | ||||||
| 8.1. Start with Beam Align. A target will appear. Using the knobs, move the beam to the center of the "bulls-eye". | ||||||
| 8.2. Next, select Aperture Align. Eliminate any shift in the window using the knobs. It is easiest to adjust so that the image shifts only in one direction (x or y). Use a magnification where you can easily see the shifting and adjust. Then move to a higher magnification and repeat. VIEW THE ALIGNMENT DEMO | ||||||
| 8.3. Click on Stig Align X and Stig Align Y. Adjust as above, but these often do not need to be changed! VIEW THE STIGMATION DEMO | ||||||
| 8.4. Close the window. | ||||||
| 9. When the area of interest is identified and the correct working distance and tilt setting are selected, lock the stage. NOTE: DO NOT CHANGE THE WORKING DISTANCE WHILE STAGE LOCK IS ON. You may need to find your location again, then refocus and stigmate. Wait a few seconds until the image stops drifting. | VIEW | |||||
| 10. Once you are satisfied with the image, check that the beam current is at least 8 ľA. If not, click the SET icon to the upper left of the voltage box. | VIEW | |||||
| VIEW | ||||||
| 14. Use the Captured Image menu to choose your image file settings. Save your images once they have been collected in a TIF format. The proper location for your image is D:\Images\users\yourname\. | VIEW | |||||