NOTE: The first user of the day must flash the tip.
| 1. Open the HV Control window by left clicking anywhere in the gray accelerating voltage area, in the top right corner of the viewing screen. | VIEW | |||||
| 2. You will need to record the values for IP1, IP2, IP3, Ie, Iflash, and Vext in the logbook. | VIEW | |||||
| 3. An HV Control dialogue box will appear. Click on Flashing. | VIEW | |||||
| 4. A dialogue box will appear asking, "Flashing Execute OK?". Click Execute. | VIEW | |||||
| 5. You will need to quickly take note of the flashing current (Ie) located in the accelerating voltage box and the value for IP1 (Iflash). | ||||||
| 6. Select an accelerating voltage and click ON. Record the value for Vext in the logbook. | VIEW | |||||