Backscatter Imaging
Backscatter imaging is primarily a compositional imaging mode, although it can be used on samples which exhibit excessive charging or on samples which provide poor SE images. The BSE detector can detect the greatest amount of signal if the stage is at zero degrees tilt. The user may need to adjust the current or spot size to increase contrast in the image.
| VIEW | ||||||
| 1.1. Raise the emission current to no more than 20 µA. This will increase the signal to noise ratio without affecting the resolution or alignment settings. | ||||||
| 1.2. Increase the spot size using the Cond Lens 1 setting in the Column Set-up window. The lens is set to a default of 5. A setting of 1 creates the largest spot size and 16 the smallest. | ||||||
| 1.3. Select the Set icon in the voltage box to re-establish the desired current. | VIEW | |||||
| 2. Before switching to BSE collection, perform alignments and stigmation in SE mode. It is important to keep the beam current higher while in BSE mode. | ||||||
| VIEW | ||||||
| 4. The default settings are: | ||||||
| Emission current = 10 µA Condenser Lens 1 = 5.0 to 8.0 Aperture = 50 µm |
||||||